| Parameters | Units | About | Contact | |||||||||
| + | 3 reference(s) hidden. |
| Emerson Process Management | Rosemount Analytical 5081-A-FF | 0206 | Transducer | |
| Rosemount Analytical 5081-C/T-FF | 0206 | Transducer | ||
| Rosemount Analytical 5081-pH/ORP-FF | 0204 | Transducer |
| Description: | Faults |
| Size: | 2 |
| Category: | Vendor |
| Access: | Engineer |
| Type: | bit Array |
| Help: | Device and sensor failures |
| Enumerations: | 11 – Broken pH Glass 12 – pH Glass Z High 13 – RTD Resistance Overrange 14 – Rtd Open 15 – A/D Converter Failure |
| + | 2 reference(s) hidden. |
| Emerson Process Management | Rosemount Analytical 5081-A-FF | 0101 | Transducer | |
| Rosemount Analytical 5081-pH/ORP-FF | 0101 | Transducer |
| Description: | Faults (Additional Transmitter Status Faults) |
| Size: | 2 |
| Category: | Vendor |
| Access: | View Only |
| Type: | bit Array |
| Help: | Additional Transmitter Status (Faults) |
| Enumerations: | 0 – Not Used 1 – Not Used 2 – Not Used 3 – Not Used 4 – Not Used 5 – Not Used 6 – Amp Fail 7 – Not Used 8 – ROM Fail 9 – CPU RAM/EEPROM Fail 10 – Factory Calibration Fail 11 – Glass Fail 12 – Rtd Open 13 – Sense Open 14 – Temperature Too Low 15 – Temperature Too High |