| Emerson Process Management | Rosemount Analytical 5081-A-FF | 0206 | Transducer | |
| Rosemount Analytical 5081-C/T-FF | 0206 | Transducer | ||
| Rosemount Analytical 5081-pH/ORP-FF | 0204 | Transducer |
| Description: | Warnings |
| Size: | 2 |
| Category: | Vendor |
| Access: | Engineer |
| Type: | bit Array |
| Help: | Device and sensor warnings |
| Enumerations: | 2 – No pH Soln Ground 3 – pH Voltage Too High 4 – Need Zero Cal 5 – Sensor Current Low 6 – Sensor Current High 7 – RTD Sense Line Open 8 – Temperature Too Low 9 – Temperature Too High 10 – PV > Display Limit 11 – Need Factory Cal 12 – Ground > 10% Off 13 – EEPROM Write Error 14 – EEPROM Buffer Overflow 15 – EEPROM Checksum Error |
| + | 2 reference(s) hidden. |
| Emerson Process Management | Rosemount Analytical 5081-A-FF | 0101 | Transducer | |
| Rosemount Analytical 5081-pH/ORP-FF | 0101 | Transducer |
| Description: | Warnings (Additional Transmitter Status Warnning) |
| Size: | 2 |
| Category: | Vendor |
| Access: | View Only |
| Type: | bit Array |
| Help: | Additional Transmitter Status (Warnings) |
| Enumerations: | 0 – Not Used 1 – Not Used 2 – Not Used 3 – Not Used 4 – Not Used 5 – Not Used 6 – ADC Overflow 7 – Bad Rtd 8 – Need Zero Cal 9 – Zero too big 10 – A/D Converter 11 – Bad Amperometric Sensor 12 – In Too Big 13 – Bad Ground 14 – PH input 15 – Over Range |