Foundation Fieldbus Parameter Search (βeta)

"Taking the frustration out of foundation"

Parameters Units About Contact
 
Parameter:
BLOCK_ERROR_CODE SELF_CALIB_ERROR_CODE

+2 reference(s) hidden.
Berthold Technologies GmbH & Co. KGUniprobe LB490/10101Resource
0202Resource
Parent:ERROR_LOG_DATA
Category:Vendor
Access:View Only
Type:Enumeration
Help:Error code number for identification of the error.
Enumerations:0 – 0 Fatal Error RAM
1 – 1 Fatal Error ROM
2 – 2 Fatal Error EEPROM
3 – 3 Error CPU clock
4 – 4 Error Power Fail
5 – 5 Error No Pulse Rate
6 – 6 Error Wrong HV
7 – 7 Error HV Deviation
8 – 8 Warning Change Multiplier
9 – 9 Error Detector temperature out of range
10 – 10 Error Analog Frontend
11 – 11 Error control count rate
12 – 12 Undefined Alarm ID
13 – 13 Undefined Alarm ID
14 – 14 Undefined Alarm ID
15 – 15 Undefined Alarm ID
16 – 16 Undefined Alarm ID
17 – 17 Error RTCbackup / Battery
18 – 18 Error RTC error
19 – 19 Error RS485 Communication
20 – 20 Warning cps < Min
21 – 21 Warning cps > Max
22 – 22 Warning Rad.Interference
23 – 23 Error Current Out
24 – 24 Undefined Alarm ID
25 – 25 Undefined Alarm ID
26 – 26 Undefined Alarm ID
27 – 27 Undefined Alarm ID
28 – No error
29 – 29 Current Input 1 Product temperature out of range
30 – 30 Current Input 2 Volumetric flow out of range
31 – 31 Pt100 Temperature out of range