Foundation Fieldbus Parameter Search (βeta)

"Taking the frustration out of foundation"

Parameters Units About Contact
 
Parameter:
+1 reference(s) hidden.
Berthold Technologies GmbH & Co. KGUniprobe LB490/10202LevelTransducer
Minimum Range:0
Maximum Range:999999
Category:Vendor
Access:Engineer
Type:32 bit Float
Help:Service function: Enter a pulse rate and check if the measuring reading shows the expected value. The Test Mode is enabled if you enter a pulse rate >0. During Test Mode the failure relay indicates an alarm. ATTENTION: Reset the test pulse rate to 0 as soon as the test is over to continue in the normal measurement mode.