| Emerson Process Management | Rosemount Analytical 1500XA | 0101 | GC | Transducer |
| Description: | Miscellaneous Status |
| Parent: | DETAILED_STATUS |
| Size: | 4 |
| Category: | Vendor |
| Access: | View Only |
| Type: | bit Array |
| Help: | Miscellaneous Detailed Status |
| Enumerations: | 8 – LOSS OF PURGE 9 – NO SAMPLE FLOW 2 10 – NO SAMPLE FLOW 1 11 – INCOMPLETE INITIALIZATION 12 – GC IN MAINTENANCE MODE 13 – ANALYSIS 2 VALIDATION FAILURE 14 – ANALYSIS 1 VALIDATION FAILURE 15 – GC DEVICE COMM FAILURE 16 – FID OVER TEMPERATURE 17 – FID FLAME OUT 18 – PREAMP 2 DET 2 EXCESS SCALING FACTOR DEVIATION 19 – PREAMP 2 DET 1 EXCESS SCALING FACTOR DEVIATION 20 – PREAMP 1 DET 2 EXCESS SCALING FACTOR DEVIATION 21 – PREAMP 1 DET 1 EXCESS SCALING FACTOR DEVIATION 22 – BASE IO BOARD COMM FAILURE 23 – PREAMP BOARD 2 COMM FAILURE 24 – PREAMP BOARD 1 COMM FAILURE 25 – HEATER SOLENOID DRIVER BOARD 2 COMM FAILURE 26 – HEATER SOLENOID DRIVER BOARD 1 COMM FAILURE 27 – WARM START FAILURE 28 – Analysis 2 CAL FAILURE 29 – Analysis 1 CAL FAILURE 30 – Analysis 2 IDLE 31 – Analysis 1 IDLE |